Extech PH90 Waterproof pH Meter Instruction Manual

Model: PH90

1. Koniec produktuview

The Extech PH90 Waterproof pH Meter is a precision instrument designed for accurate pH measurement across various applications, including water quality testing, hydroponics, aquaculture, swimming pools, beverage production, and laboratory use. This device features a robust waterproof housing, a replaceable electrode, 3-point automatic calibration, and auto buffer recognition for ease of use and reliable results. Its compact design makes it suitable for both field and professional environments.

Extech PH90 Waterproof pH Meter

Figure 1: Extech PH90 Waterproof pH Meter. This image shows the compact, handheld pH meter with its digital display and control buttons, ready for use.

2. Główne cechy

3. Konfiguracja i przygotowanie

3.1. Instalacja baterii

The PH90 meter requires two CR2032 Lithium Metal batteries (included). To install or replace batteries:

  1. Unscrew the battery compartment cap located at the top of the meter.
  2. Insert the two CR2032 batteries, ensuring correct polarity (+/-).
  3. Securely screw the battery compartment cap back into place.

3.2. Electrode Preparation

Before first use, and after storage, prepare the electrode:

  1. Remove the protective cap from the electrode tip. The cap contains a moist sponge to keep the electrode hydrated.
  2. Rinse the electrode tip thoroughly with distilled or deionized water. Do not wipe the electrode, as this can create static charges.
  3. Allow any excess water to drip off.

3.3. Kalibracja wstępna

For optimal accuracy, perform a 3-point calibration before initial use and periodically thereafter. Calibration buffers (pH 4, 7, and 10) are sold separately.

  1. Turn on the meter by pressing the WŁ./WYŁ. przycisk.
  2. Immerse the electrode into a pH 7.00 buffer solution.
  3. Naciśnij i przytrzymaj CAL button until "CAL" appears on the display. The meter will automatically recognize the buffer and calibrate.
  4. Przepłucz elektrodę wodą destylowaną.
  5. Repeat the process with pH 4.00 and pH 10.00 buffer solutions. The meter will automatically detect and calibrate to these points.
  6. Po zakończeniu kalibracji miernik powróci do trybu pomiaru.
Extech PH90 Meter Display and Buttons

Figure 2: Close-up of the Extech PH90 display and control buttons, including ON/OFF, MODE, and CAL buttons.

4. Instrukcja obsługi

4.1. Wykonywanie pomiaru

  1. Sprawdź, czy miernik jest skalibrowany.
  2. Rinse the electrode with distilled water and gently shake off excess.
  3. Zanurz końcówkę elektrody wampRozwiązanie polega na całkowitym zanurzeniu czujnika.
  4. Gently stir the solution to ensure homogeneity and allow the reading to stabilize. This may take up to a minute, especially if the sample temperature differs significantly from the meter's previous environment.
  5. Read the pH value displayed on the LCD screen. The temperature will also be displayed.
  6. After measurement, rinse the electrode thoroughly with distilled water.

4.2. Funkcja wstrzymania

Aby zamrozić aktualny odczyt na wyświetlaczu, naciśnij przycisk TRZYMAĆ button (if available, typically integrated with MODE or a separate button). Press it again to release the hold and return to live measurement.

4.3. Automatyczna kompensacja temperatury (ATC)

The Extech PH90 features Automatic Temperature Compensation, which adjusts pH readings based on the sample's temperature, ensuring accuracy across varying temperatures. The temperature is displayed alongside the pH value.

5. Konserwacja

5.1. Pielęgnacja i przechowywanie elektrod

Proper electrode care is crucial for maintaining accuracy and extending the life of your meter:

Extech PH90 Electrode Tip

Figure 3: Close-up of the Extech PH90's replaceable electrode tip, which requires proper care and storage.

5.2. Wymiana elektrod

The PH90 features a replaceable electrode. If the electrode performance degrades significantly despite proper cleaning and calibration, it may need replacement. Refer to the Extech webodwiedź witrynę lub skontaktuj się z obsługą klienta, aby uzyskać informacje o zakupieasing replacement electrodes (Part PH95).

5.3. Wymiana baterii

When the battery indicator appears on the display or the meter fails to power on, replace the two CR2032 Lithium Metal batteries as described in Section 3.1.

6. Rozwiązywanie Problemów

ProblemMożliwa przyczynaRozwiązanie
Niedokładne lub niestabilne odczyty
  • Electrode not calibrated
  • Dirty or dry electrode
  • Temperature not equilibrated
  • Expired buffer solutions
  • Uszkodzona elektroda
  • Wykonaj kalibrację 3-punktową.
  • Clean the electrode and ensure it is properly stored.
  • Allow sufficient time for the meter's temperature sensor to equilibrate with the sample.
  • Use fresh calibration buffer solutions.
  • Wymień elektrodę.
Miernik się nie włącza
  • Rozładowane baterie
  • Nieprawidłowa instalacja baterii
  • Wymień baterie na nowe ogniwa CR2032.
  • Sprawdź biegunowość baterii.
"CAL" error during calibration
  • Użyto niewłaściwego roztworu buforowego
  • Contaminated buffer solution
  • Electrode issue
  • Ensure correct pH buffer (4, 7, or 10) is used.
  • Use fresh, uncontaminated buffer solutions.
  • Clean or replace the electrode.

7. Specyfikacje

ParametrWartość
Numer modeluPH90
Zakres pH0 do 14 pH
Dokładność pH±0.01 pH
Zakres temperatur(Not explicitly stated, but integrated for ATC)
Źródło zasilania2 x CR2032 Lithium Metal batteries
Waga przedmiotu2.88 uncji (około 81.6 gramów)
Wymiary produktu6.7 x 1.4 x 1.4 cala (ok. 17 x 3.6 x 3.6 cm)
Zawarte komponentyFlat surface pH electrode, cap, batteries (pH buffers sold separately)
ProducentWydobycie

8. Gwarancja i wsparcie

For warranty information, technical support, or to purchase replacement parts and accessories, please visit the official Extech website or contact Extech customer service. You can also find more information and products at the Sklep Extech na Amazon.