Extech FL700

Instrukcja obsługi miernika fluoru Extech FL700

Model: FL700

1. Wprowadzenie

The Extech FL700 Fluoride Meter is a precision instrument designed for the rapid and accurate measurement of fluoride ions in various aqueous samples, including drinking water. This device integrates a sensing electrode, measurement electronics, and a digital display into a single, portable unit, making it suitable for both field and laboratory applications.

The FL700 provides direct readings of total fluoride concentrations ranging from 9.99 ppm or mg/L down to 0.1 ppm. It features automatic temperature compensation to ensure measurement accuracy across varying environmental conditions.

2. Informacje dotyczące bezpieczeństwa

Please read and understand all safety precautions before operating the Extech FL700 Fluoride Meter. Failure to follow these instructions may result in injury or damage to the instrument.

  • Always handle the fluoride electrode with care. Avoid touching the sensing element directly.
  • When working with TISAB reagent tablets or any chemical solutions, wear appropriate personal protective equipment, including gloves and eye protection.
  • Do not immerse the entire meter in liquid. The meter is designed for electrode immersion only.
  • Store the meter and reagents in a cool, dry place, away from direct sunlight and extreme temperatures.
  • Przechowywać w miejscu niedostępnym dla dzieci.

3. Koniec produktuview

The Extech FL700 Fluoride Meter is a compact and user-friendly device. Key components and features are outlined below:

Extech FL700 Fluoride Meter front view
Przód view of the Extech FL700 Fluoride Meter, showing the display and control buttons.
Hand holding Extech FL700 Fluoride Meter
The Extech FL700 Fluoride Meter held in hand, demonstrating its ergonomic design and clear display.
Close-up of Extech FL700 electrode tip
Z bliska view of the fluoride electrode tip, which is integrated into the meter.
Extech FL700 Fluoride Meter in use with sample
The Extech FL700 Fluoride Meter being used to measure a liquid sample, showing the meter in its stand.
Extech FL700 Fluoride Meter and included accessories
All components included with the Extech FL700: the meter, sensor cap, TISAB reagent tablets, and neckstrap.

Główne cechy:

  • Kalibracja dwupunktowa: Ensures accurate measurements across the fluoride range.
  • Pamięć masowa: Stores up to 25 labeled readings for later review.
  • Field or lab testing: Designed for versatile use in various environments.
  • Szybka odpowiedź: Provides fluoride measurements in less than 1 minute.
  • Large dual LCD: Simultaneously displays ppm (or mg/L) and temperature with a bargraph.

4. Konfiguracja

4.1. Instalacja baterii

The Extech FL700 requires four 3V button batteries (included). To install or replace batteries:

  1. Twist off the battery compartment cap located at the top of the meter.
  2. Insert the four 3V button batteries, ensuring correct polarity as indicated inside the compartment.
  3. Replace the battery compartment cap and twist to secure.

4.2. Wstępne przygotowanie

Before first use or after prolonged storage, prepare the electrode:

  1. Remove the sensor cap from the electrode tip.
  2. Rinse the electrode tip with distilled water.
  3. Condition the electrode by immersing it in a fluoride standard solution (e.g., 1 ppm or 10 ppm) for approximately 15-30 minutes.

5. Instrukcja obsługi

5.1. Włączanie/wyłączanie zasilania

Naciśnij WŁ./WYŁ. przycisk włączania i wyłączania miernika.

5.2. Calibration (Two-Point)

Accurate calibration is crucial for reliable measurements. The FL700 supports two-point calibration.

  1. Prepare two fluoride standard solutions, typically one low concentration (e.g., 1 ppm) and one high concentration (e.g., 10 ppm). Ensure the solutions are at a stable temperature.
  2. Opłucz elektrodę wodą destylowaną i osusz.
  3. Immerse the electrode into the first standard solution.
  4. Naciśnij i przytrzymaj KAL/PRZYWRÓĆ button until 'CAL' appears on the display. The meter will automatically recognize the standard and calibrate. Wait for the reading to stabilize.
  5. Rinse the electrode again and immerse it into the second standard solution.
  6. Naciśnij i przytrzymaj KAL/PRZYWRÓĆ button again. The meter will complete the two-point calibration.
  7. The meter is now calibrated and ready for measurement.

5.3. Wykonywanie pomiaru

To measure fluoride concentration in a sampna:

  1. Sprawdź, czy miernik jest skalibrowany.
  2. Opłucz elektrodę wodą destylowaną i osusz.
  3. Add TISAB (Total Ionic Strength Adjustment Buffer) reagent to your sample according to the reagent instructions. TISAB ensures that the fluoride ions are in a measurable form and minimizes interference.
  4. Immerse the electrode tip into the prepared sample solution. Ensure the tip is fully submerged.
  5. Allow the reading to stabilize on the display. The large dual LCD will show the fluoride concentration in ppm (or mg/L) and the temperature.
  6. Aby zachować aktualny odczyt, naciśnij przycisk TRYB/WSTRZYMANIE przycisk. Naciśnij ponownie, aby zwolnić.

5.4. Recalling Stored Readings

The meter can store up to 25 labeled readings.

  • To recall stored readings, press the KAL/PRZYWRÓĆ button briefly when not in calibration mode.
  • Użyj TRYB/WSTRZYMANIE button to scroll through the stored readings.

6. Konserwacja

6.1. Czyszczenie elektrod

Regular cleaning of the electrode ensures optimal performance.

  • After each use, rinse the electrode thoroughly with distilled water.
  • If the electrode becomes fouled or readings become erratic, clean it with a mild detergent solution or a specialized electrode cleaning solution. Rinse thoroughly with distilled water afterward.

6.2. Przechowywanie

Proper storage extends the life of the electrode and meter.

  • Always replace the sensor cap on the electrode tip when not in use. The cap should contain a small amount of electrode storage solution or a fluoride standard solution to keep the membrane hydrated. Do not store dry.
  • Przechowuj miernik w chłodnym i suchym miejscu.

6.3. Wymiana baterii

Replace batteries when the low battery indicator appears on the display or when the meter fails to power on. Refer to Section 4.1 for battery installation instructions.

7. Rozwiązywanie Problemów

ProblemMożliwa przyczynaRozwiązanie
Miernik się nie włączaWyczerpane lub nieprawidłowo zainstalowane baterieSprawdź biegunowość baterii; wymień baterie.
Erratic or unstable readingsDirty electrode; air bubbles on electrode; insufficient sample volume; expired or incorrect TISAB.Clean electrode; gently tap electrode to remove bubbles; ensure electrode is fully submerged; check TISAB.
Meter fails to calibrateIncorrect standard solutions; dirty electrode; electrode not conditioned; damaged electrode.Verify standard concentrations; clean and condition electrode; replace electrode if damaged.
Wskaźnik niskiego poziomu naładowania bateriiBaterie są słabe.Replace all four 3V button batteries.

8. Specyfikacje

SpecyfikacjaSzczegół
ProducentWydobycie
Numer częściFL700
Waga przedmiotu7.7 uncji
Wymiary produktu6.8 x 1.6 x 1.4 cala
Numer modelu przedmiotuFL700
BaterieWymagane 4 baterie litowo-metalowe (w zestawie)
StylFL700
Ilość w opakowaniu przedmiotu1
Liczba sztuk1
Zawarte komponentyJednostka, instrukcja obsługi
Typ ogniwa bateriiAlkaliczny
Pierwsza dostępna data15 lipca 2008 r.

9. Informacje o gwarancji

The Extech FL700 Fluoride Meter comes with a full one-year warranty on parts and workmanship, unless otherwise specified. This warranty does not cover defects resulting from user actions such as misuse, improper wiring, operation outside of specified parameters, improper maintenance or repair, or unauthorized modification.

10. Wsparcie

For technical support, service, or additional information, please visit the official Extech webna stronie internetowej lub skontaktuj się z działem obsługi klienta. Więcej informacji i produktów znajdziesz również na stronie Sklep Extech na Amazon.

Powiązane dokumenty - FL700

Przedview Instrukcja obsługi miernika wilgotności, temperatury, przepływu powietrza i światła EXTECH 45170 4 w 1
Instrukcja obsługi miernika EXTECH 45170, 4-w-1 mierzącego wilgotność, temperaturę, przepływ powietrza i oświetlenie. Zawiera informacje dotyczące obsługi, specyfikacji i gwarancji.
Przedview Instrukcja obsługi promieni UV-AB Extech UV505
Kompletna instrukcja obsługi Misuratore di Radiazione UV-AB Extech UV505. Uwzględnij instrukcje operacyjne, specyficzne techniki, informacje o sicurezza i manutenzione.
Przedview Instrukcja obsługi miernika światła UV-AB Extech UV505
Instrukcja obsługi światłomierza Extech UV505 UV-AB. Niniejszy dokument zawiera szczegółowe informacje na temat funkcji urządzenia, jego obsługi, specyfikacji, środków ostrożności i konserwacji w celu pomiaru promieniowania UVA i UVB.
Przedview Extech EX810 1000 Amp Klemmeter z termometrem na podczerwień Gebruikershandleiding
Gedetailleerde gebruikershandleiding voor de Extech EX810 1000 Amp klemmeter z ingebouwde termometr na podczerwień. Bevat instructies, veiligheidsrichtlijnen en technische specaties voor nauwkeurige metingen van spanning, stroom, weerstand, capaciteit, częstie en temperatuur.
Przedview Instrukcja obsługi serii Extech EX650: Cyfrowy wyświetlacz Cl 600A True RMSamp Metry
Kompleksowy podręcznik użytkownika dla serii Extech EX650 600A True RMS cyfrowych kl.amp Liczniki, obejmujące modele EX650 i EX655. Dowiedz się więcej o obsłudze, funkcjach, bezpieczeństwie i specyfikacjach dotyczących dokładnego pomiaru objętości prądu przemiennego/stałego.tage, prąd, rezystancja i inne pomiary.
Przedview Instrukcja obsługi kieszonkowego multimetru Extech DM100
Instrukcja obsługi kieszonkowego multimetru automatycznego z automatycznym zakresem DM100 firmy Extech o pojemności 4000 jednostek, obejmująca zagadnienia dotyczące bezpieczeństwa, danych technicznych, obsługi, funkcji i konserwacji.