ZGOOD 18L Autoclave Automatic Sterilizer

ZGOOD 18L Tabletop Class N High Pressure Steam Sterilizer User Manual

Model: 18L Autoclave Automatic Sterilizer

1. Koniec produktuview

The ZGOOD 18L Tabletop Class N High Pressure Steam Sterilizer is an automatic sterilizer designed for efficient and reliable sterilization of instruments and materials. It features a drying function and real-time monitoring of pressure and temperature, offering two preset programs and adjustable duration for specific sterilization requirements.

Główne cechy:

  • Efficient Automatic Sterilization: Operates at 1700W with easy operation, real-time pressure and temperature monitoring, and adjustable sterilization duration beyond preset programs (121°C/50 min, 134°C/40 min).
  • Przejrzysty wyświetlacz LED: Three LED displays for pressure, temperature, and dry status, providing clear process and time information.
  • Automatic Pressure Release & Drying Function: Automatically releases internal pressure and discharges water/steam after completion. The drying function reduces the drying cycle and improves drying effectiveness for internal components and chambers.
  • Durable Stainless Steel Chamber: Constructed from stainless steel with a double-layer stainless steel frame and an aluminum instrument tray for maximum capacity.
  • Wielorakie zabezpieczenia: Includes temperature and pressure overload protection, pressure safety lock (prevents door opening until pressure is zero), and safety door locks (ensures operation only when door is correctly locked).

2. Instrukcje bezpieczeństwa

Read all instructions carefully before operating the sterilizer. Failure to follow safety guidelines may result in injury or damage to the equipment.

  • Bezpieczeństwo elektryczne: Ensure the electrical supply matches the unit's requirements (110 ~ 230 VAC). Do not operate with damaged cords or plugs.
  • High Pressure and Temperature: The sterilizer operates under high pressure and temperature. Never attempt to open the door during operation or when the pressure gauge indicates internal pressure. The pressure safety lock prevents opening until pressure is zero.
  • Ryzyko poparzenia: Surfaces and internal components will be extremely hot after a cycle. Use appropriate protective gear (e.g., heat-resistant gloves) when handling sterilized items or performing maintenance.
  • Jakość wody: Use only distilled water for operation to prevent mineral buildup and damage to the unit.
  • Załadunek: Do not overload the chamber. Ensure items are arranged to allow proper steam circulation. Disinfection equipment should not be excessive and should not be stacked.
  • Wentylacja: Podczas pracy urządzenia należy zapewnić odpowiednią wentylację.
  • Wyłączenie awaryjne: In case of malfunction or emergency, immediately switch off the main power.

3. Zawartość opakowania

Sprawdź, czy po rozpakowaniu znajdują się wszystkie elementy:

  • Main Sterilizer Unit
  • Instrument Racks
  • Taca na instrumenty
  • Przewód zasilający
  • Instrukcja obsługi
  • Measuring Cup (for water)
  • Tray Removal Tool
ZGOOD 18L Autoclave Sterilizer with door open, showing the interior chamber and racks.

Image: The ZGOOD 18L Autoclave Sterilizer with its door open, revealing the stainless steel interior and instrument racks.

4. Składniki produktu

Familiarize yourself with the main components of the sterilizer:

Diagram of ZGOOD 18L Autoclave Sterilizer with labeled components: Pressure Gauge, LED Display, Door Handle, Exhaust Valve, Main Switch, Safety Valve, Power Jack.

Image: Labeled diagram illustrating the key components of the ZGOOD 18L Autoclave Sterilizer.

Sterilizer Components
NIE.Część
1Ciśnieniomierz
2Wyświetlacz LED
3Klamka do drzwi
4Zawór wydechowy
5Wyłącznik główny
6Zawór bezpieczeństwa
7Gniazdo zasilania

5. Konfiguracja

  1. Rozpakowywanie: Carefully remove the sterilizer and all accessories from the packaging. Retain packaging for future transport or storage.
  2. Umieszczenie: Place the unit on a stable, level, heat-resistant surface with adequate clearance for ventilation. Ensure it is away from heat sources and direct sunlight.
  3. Podłączenie zasilania: Connect the power cord to the sterilizer's power jack and then to a grounded electrical outlet that matches the specified voltage (110 ~ 230 VAC).
  4. Add Distilled Water: Open the door by rotating the door handle counter-clockwise and pulling the door open. Using the provided measuring cup, add the appropriate amount of distilled water to the chamber. Important: Only use distilled water.

Video: Demonstrates unpacking, adding water, and loading instruments into the sterilizer. This video shows the initial setup and preparation steps for sterilization.

6. Instrukcja obsługi

  1. Załaduj instrumenty: Place instruments and materials to be sterilized onto the instrument racks. Ensure items are not excessively stacked to allow proper steam penetration. Wrapped items can be placed in sterilized cloth.
  2. Zamknij i zabezpiecz drzwi: Close the sterilizer door and rotate the door handle clockwise to securely tighten it. Ensure the door is fully sealed.
  3. Włączanie: Flip the main switch to the "ON" position. The LED display will illuminate.
  4. Select Sterilization Program: Use the control buttons on the LED display to select the desired temperature (121°C or 134°C) and sterilization time. The unit has preset programs (121°C/50 min; 134°C/40 min) but allows for adjustment.
  5. Rozpocznij sterylizację: Press the "Start" button to begin the sterilization cycle. The display will show the current temperature and pressure.
  6. Monitoruj postęp: The unit will automatically monitor temperature and pressure. The display will show "End" when the cycle is complete.
  7. Funkcja suszenia: After sterilization, the unit will automatically release pressure and initiate the drying function to ensure items are dry.
  8. Otwarte drzwi: Once the display shows "End" and the pressure gauge indicates zero pressure, carefully rotate the door handle counter-clockwise and open the door. Use the tray removal tool to prevent burns when retrieving hot items.

Video: Illustrates the operational steps of the sterilizer, including powering on, setting temperature, starting disinfection, and safely opening the door after the cycle.

ZGOOD 18L Autoclave Sterilizer with various items like dental tools, wrapped packages, and face masks, indicating suitable items for sterilization.

Obraz: Byłyamples of items suitable for sterilization in the autoclave, including exposed metal tools, wrapped items, and cotton fabrics.

7. Konserwacja

Regular maintenance ensures optimal performance and longevity of your sterilizer.

  • Codzienne sprzątanie: After each use, wipe the interior of the chamber and the door gasket with a soft, damp szmatką. Upewnij się, że nie pozostały żadne pozostałości.
  • Odprowadzanie wody: Regularly drain any remaining water from the reservoir. Refer to the specific drainage port location in the product diagram.
  • Kontrola uszczelki: Periodically inspect the door gasket for any signs of wear, cracks, or damage. A damaged gasket can lead to steam leaks and improper sterilization. Replace if necessary.
  • Czyszczenie zewnętrzne: Keep the exterior of the unit clean and free of dust. Use a mild detergent if needed, but avoid abrasive cleaners.
  • Składowanie: Jeśli urządzenie będzie przechowywane przez dłuższy okres czasu, należy upewnić się, że jest czyste, suche i przechowywane w chłodnym, suchym miejscu.

8. Rozwiązywanie Problemów

W tej sekcji omówiono typowe problemy, które mogą wystąpić. W przypadku problemów, których tutaj nie wymieniono, skontaktuj się z obsługą klienta.

Typowe problemy z rozwiązywaniem problemów
ProblemMożliwa przyczynaRozwiązanie
Urządzenie nie włącza sięNo power supply; Main switch off; Blown fuseCheck power cord connection; Ensure main switch is ON; Check circuit breaker or replace fuse (if accessible and safe).
Steam leaks from doorDoor not properly sealed; Damaged door gasketEnsure door handle is fully tightened; Inspect and replace door gasket if damaged.
Sterilization cycle does not startDoor not closed properly; Insufficient water; Program not selectedVerify door is securely closed; Check water level and add distilled water if needed; Select a sterilization program.
Items are wet after drying cycleOverloaded chamber; Insufficient drying time; Items too denseReduce load size; Increase drying time if adjustable; Arrange items to allow better air circulation.

9. Specyfikacje

Specyfikacja produktu
FunkcjaSzczegół
Nazwa produktu18L Autoclave Sterilizer
Zasilanie elektryczne110 ~ 230 V AC
Chamber Dimensions (Diameter/Depth)Ř247 × 340 mm
Moc1700 W
Temperatura pracy121°C, 134°C
Wymiary całkowite (długość x szerokość x wysokość)Wymiary: 568 × 419 × 336 mm
Masa netto25.9 KG
Working Time (Preset)121°C/50 min; 134°C/40 min
ProducentZDOBRY
ASINB0GS5J185H

10. Gwarancja i wsparcie

For warranty information, technical support, or service inquiries, please refer to the contact details provided with your purchase documentation or visit the official ZGOOD webZachowaj paragon jako dowód zakupu na wypadek reklamacji.

© 2026 ZGOOD. All rights reserved.

Ask a question about this manual

Ask about setup, troubleshooting, compatibility, parts, safety, or missing instructions. Manuals+ will review the question and use this page’s manual context to help answer it.

After a successful submission, we securely hash the supplied account or form email before sending it to Microsoft Advertising for conversion attribution. Privacy details.