YM101

Instrukcja obsługi profesjonalnego pulsoksymetru palcowego YM101

Model: YM101 | Brand: Generic

1. Wprowadzenie

This manual provides instructions for the Generic Professional Finger Pulse Oximeter YM101, a portable device designed for non-invasive spot-checking of blood oxygen saturation (SpO2) and pulse rate (PR) at home or in professional settings. It features an OLED display for clear readings and is suitable for various users, including children and the elderly. The device is CE approved, ensuring accurate and reliable measurements.

2. Ważne informacje dotyczące bezpieczeństwa

Please read all safety information before using the device. Failure to follow these instructions may result in inaccurate readings or device damage.

  • The pulse oximeter is intended for spot-checking and not for continuous monitoring.
  • This device is not intended to diagnose, treat, cure, or prevent any disease. Consult a medical professional for any health concerns.
  • Nie używaj urządzenia w środowisku MRI lub CT.
  • Trzymaj urządzenie z dala od wody i wilgoci.
  • Przed dokonaniem pomiaru należy upewnić się, że palec jest czysty i suchy.
  • Remain still during measurement to ensure accuracy. Movement can affect readings.
  • Do not use the device on the same arm as a blood pressure cuff or arterial catheter.
  • Zutylizuj baterie zgodnie z lokalnymi przepisami.

3. Zawartość opakowania

  • 1 x Generic Professional Finger Pulse Oximeter YM101
  • 1 x Smycz
  • 1 x Instrukcja obsługi (wielojęzyczna)

Note: 2 AAA 1.5V batteries are required but not included in the package.

4. Urządzenie ponadview

The Generic Professional Finger Pulse Oximeter YM101 is a compact and lightweight device designed for ease of use. It features a bright OLED display for clear visibility of readings.

Generic Professional Finger Pulse Oximeter YM101 with OLED display showing SpO2 and PR bpm readings
Rysunek 1: Przód view of the YM101 Pulse Oximeter, displaying blood oxygen saturation (SpO2) and pulse rate (PR bpm) on its OLED screen. The device has a single power button below the display.

The device body is designed with special lighting technology to ensure accurate values even in strong ambient light conditions. The OLED screen displays blood oxygen saturation, pulse rate, pulse waveform, and a histogram.

5. Instalacja baterii

The YM101 Pulse Oximeter requires two AAA 1.5V batteries for operation. Ensure correct polarity during installation.

  1. Otwórz pokrywę komory baterii znajdującą się z tyłu urządzenia.
  2. Insert two AAA batteries according to the polarity indicators (+ and -) inside the compartment.
  3. Dokładnie zamknij pokrywę komory baterii.
Open battery compartment of the YM101 Pulse Oximeter, showing where to insert two AAA batteries
Rysunek 2: View of the open battery compartment, illustrating the correct placement for two AAA batteries. The battery cover is shown detached.

The device features low power consumption and will display a warning when battery voltage is low. It automatically shuts off after 8 seconds of inactivity to conserve battery life, providing up to 24 hours of continuous use.

6. Instrukcja obsługi

Aby wykonać pomiar, wykonaj następujące czynności:

  1. Upewnij się, że baterie są prawidłowo zainstalowane.
  2. Otwórz klamp pulsoksymetru.
  3. Insert one of your fingers (preferably the index finger) into the rubber opening of the oximeter, ensuring your fingertip is placed correctly.
  4. Zwolnij clamp.
  5. Press the power button once on the front panel.
  6. Keep your hand still and do not move your finger during the measurement.
  7. Your SpO2 and pulse rate readings will appear on the display within approximately 8 seconds.
Person using the YM101 Pulse Oximeter on their finger
Rysunek 3: A user demonstrating the correct placement of the finger into the pulse oximeter for measurement.

7. Zrozumienie swoich odczytów

The OLED display provides clear and comprehensive information:

  • SpO2 (%): Blood Oxygen Saturation level. This indicates the percentage of hemoglobin in your blood that is saturated with oxygen.
  • PR bpm: Pulse Rate in beats per minute. This is the number of times your heart beats per minute.
  • Pulse Waveform/Bar Graph: A visual representation of your pulse strength.
YM101 Pulse Oximeter display showing multi-direction viewing opcje
Rysunek 4: The multi-directional display feature of the YM101 Pulse Oximeter, illustrating how readings can be viewed in various orientations for user convenience.

The device features a rotatable multi-directional display with 4 directions and 6 modes, allowing you to view your results in any orientation.

8. Funkcja alarmu

The YM101 Pulse Oximeter is equipped with an alarm function to alert you if your measured values fall outside predefined limits. This feature helps you to be immediately aware of any significant changes in your SpO2 or pulse rate.

YM101 Pulse Oximeter display showing an alarm icon and 'Beep~' text
Rysunek 5: The display of the pulse oximeter indicating an active alarm with a bell icon and 'Beep~' text, signifying that a measured value has exceeded or fallen below a set threshold.

Refer to the full manual for instructions on how to set or adjust alarm thresholds, if applicable.

9. Konserwacja i pielęgnacja

  • Czyszczenie: Wyczyść powierzchnię pulsoksymetru miękką ściereczką.ampened with medical alcohol (70% ethanol or isopropanol). Do not immerse the device in water or other cleaning solutions.
  • Składowanie: Przechowuj urządzenie w chłodnym, suchym miejscu, z dala od bezpośredniego światła słonecznego i ekstremalnych temperatur. Jeśli nie będziesz go używać przez dłuższy czas, wyjmij baterie.
  • Wymiana baterii: Wymień baterie, gdy na wyświetlaczu pojawi się wskaźnik niskiego poziomu naładowania baterii.
  • Sprzedaż: Dispose of the device and its components, including batteries, according to local environmental regulations.

10. Rozwiązywanie Problemów

If you encounter issues with your pulse oximeter, refer to the table below for common problems and solutions.

ProblemMożliwa przyczynaRozwiązanie
Brak wyświetlacza po włączeniuNiski poziom naładowania baterii lub rozładowane baterie; nieprawidłowa instalacja bateriiReplace batteries; Reinstall batteries with correct polarity
Niedokładne odczytyFinger not inserted properly; Patient movement; Cold fingers; Nail polish or artificial nailsReinsert finger correctly; Remain still; Warm hands; Remove nail polish/artificial nails
"Finger Out" or similar error messageFinger not detected or improperly placedUpewnij się, że palec jest całkowicie wsunięty i spoczywa na czujniku.
Urządzenie nieoczekiwanie wyłącza sięAutomatic shutdown after 8 seconds of inactivity; Low batteryThis is normal behavior; Replace batteries if low.

Jeśli po wypróbowaniu tych rozwiązań problem nadal występuje, skontaktuj się z obsługą klienta.

11. Specyfikacje techniczne

ParametrSpecyfikacja
Typ wyświetlaczaHD Color OLED Display
Zakres pomiaru SpO270% ~ 99%
Dokładność SpO2±2% (80%~99%), ±3% (70%~79%), unspecified (<70%)
Rozdzielczość SpO2±1%
Zakres pomiaru tętna30 BPM ~ 240 BPM
Dokładność tętna±1 BPM or ±1% (whichever is higher)
Pobór mocyLess than 30-50 mA
Typ baterii2 baterie AAA 1.5 V
Żywotność bateriiDo 24 godzin (ciągłe użytkowanie)
Automatyczne wyłączanie8 seconds of no operation
Temperatura pracy5℃ ~ 40℃ (41℉ ~ 104℉)
Temperatura przechowywania-10℃ ~ 40℃ (14℉ ~ 104℉)
Wilgotność otoczenia15% ~ 80% (operating), 10% ~ 80% (storage)
Ciśnienie atmosferyczne86 kPa ~ 106 kPa
Wymiary (dł. x szer. x wys.)5.5 x 3 x 3 cm (2.17 x 1.18 x 1.18 cala)
Waga120 grama (w przybliżeniu)
ProducentYONKER

12. Gwarancja i wsparcie

This product comes with a standard manufacturer's warranty. For specific warranty details, please refer to the documentation provided with your purchase or contact your retailer.

For technical support, troubleshooting assistance, or any questions regarding the use of your Generic Professional Finger Pulse Oximeter YM101, please contact the retailer from whom you purchased the device.

Powiązane dokumenty - YM101

Przedview AEPTEK YM101 Pulse Oximeter Welcome Guide
Welcome guide for the AEPTEK YM101 Finger Pulse Oximeter. Learn how to use, install batteries, customize settings, and understand readings for SpO2, pulse rate, and perfusion index. Includes tips for accurate measurements and normal ranges.