Instrukcja obsługi pipety Gilson Pipetman F100 Model F o stałej objętości, zakres objętości 100 µl

1. Wprowadzenie i koniecview

The Gilson Pipetman F100 is a fixed-volume, air-displacement pipette designed for precise liquid handling in laboratory settings. This model is specifically calibrated for a 100µl volume range, ensuring consistent and reliable results for routine testing, clinical diagnostics, and quality control applications. Its fixed-volume mechanism eliminates the risk of volume-selection errors during standard operation.

While primarily fixed, the pipette's volume can be adjusted by qualified users in a laboratory environment to accommodate for specific fluid properties, such as dense or viscous liquids. Once reset, the new volume remains fixed and cannot be accidentally altered during use.

Gilson Pipetman F100 Fixed-Volume Pipette, showing the main body, plunger, and tip ejector, with a disposable tip next to it.

Rysunek 1: The Gilson Pipetman F100 fixed-volume pipette. This image displays the blue main body with the 'Pipetman' logo, a yellow plunger mechanism visible at the top, and a white tip ejector rod extending downwards. A clear disposable pipette tip is shown separately, ready for attachment.

2. Konfiguracja

  1. Rozpakowywanie: Carefully remove the Pipetman F100 from its packaging. Inspect the pipette for any visible damage that may have occurred during transit.
  2. Inspekcja wstępna: Ensure all components are present and securely attached. The pipette should feel solid and free of loose parts.
  3. Załącznik końcówki: Select a high-quality, compatible pipette tip. Firmly press the pipette shaft into the opening of the tip with a slight twisting motion until a secure seal is formed. Avoid excessive force, which can damage the tip cone or affect accuracy.
  4. Volume Verification (Optional): Although fixed, it is recommended to perform an initial gravimetric check with distilled water to confirm the pipette's accuracy, especially if it is new or has been stored for an extended period. Refer to standard laboratory procedures for gravimetric calibration.

3. Instrukcja obsługi

The Gilson Pipetman F100 is designed for straightforward operation due to its fixed-volume nature. Follow these steps for accurate liquid transfer:

  1. Dołącz wskazówkę: Securely attach a new, clean pipette tip to the tip cone as described in the Setup section.
  2. Wstępne zwilżanie (zalecane): For optimal accuracy, especially with volatile liquids, aspirate and dispense the liquid once or twice before transferring the actual sample. This saturates the air space in the tip with vapor, reducing evaporative loss during the actual transfer.
  3. Dążenie:
    • Depress the plunger smoothly to the first stop.
    • Immerse the tip into the liquid to a depth of 2-3 mm.
    • Slowly release the plunger to allow the liquid to be drawn into the tip. Wait 1-2 seconds after the plunger reaches its resting position to ensure the full volume is aspirated.
    • Withdraw the tip from the liquid, touching it against the edge of the reservoir to remove any excess droplets.
  4. Wydawanie:
    • Przyłóż końcówkę do wewnętrznej ściany naczynia odbiorczego pod niewielkim kątem.
    • Smoothly depress the plunger to the first stop, expelling the majority of the liquid.
    • Continue depressing the plunger to the second stop (blow-out) to ensure complete delivery of the remaining liquid.
    • Przytrzymując tłok w drugim położeniu, wyciągnij pipetę z naczynia.
  5. Tip Ejection: Press the tip ejector button to discard the used tip into an appropriate waste container. Never reuse tips.

4. Konserwacja

Proper maintenance ensures the longevity and accuracy of your Gilson Pipetman F100.

  • Codzienne sprzątanie: Przetrzyj zewnętrzną część pipety miękką ściereczką.ampened with 70% ethanol or a mild detergent solution. Avoid harsh solvents.
  • Odkażenie: If the pipette comes into contact with hazardous materials, follow established laboratory decontamination protocols. The lower part of the pipette (tip ejector and tip cone) can often be autoclaved or disassembled for thorough cleaning, consult the full manufacturer's manual for specific instructions.
  • Składowanie: Store the pipette vertically in a pipette stand to prevent contamination and damage to the tip cone. Ensure it is stored in a clean, dry environment away from direct sunlight and extreme temperatures.
  • Regularna kontrola kalibracji: Even though it's a fixed-volume pipette, periodic gravimetric calibration checks (e.g., every 3-6 months or as per laboratory SOPs) are crucial to confirm continued accuracy and precision. If the pipette consistently falls outside acceptable limits, it may require servicing or recalibration by a qualified technician.

5. Rozwiązywanie Problemów

If you encounter issues with your Pipetman F100, consider the following common problems and solutions:

ProblemMożliwa przyczynaRozwiązanie
Inaccurate or inconsistent dispensing
  • Loose or incompatible tip
  • Damaged tip cone
  • Contamination inside the pipette
  • Improper pipetting technique (e.g., too fast aspiration/dispensing)
  • Temperature differences between pipette, liquid, and environment
  • Ensure tip is securely attached and compatible.
  • Inspect tip cone for cracks or damage; replace if necessary.
  • Perform thorough cleaning and decontamination.
  • Review operating instructions for correct technique.
  • Allow pipette and liquids to equilibrate to room temperature.
  • Perform a gravimetric calibration check.
Płyn kapie z końcówki
  • Loose or incompatible tip
  • Damaged tip cone or seal
  • Contamination affecting seal
  • Re-attach tip firmly or try a new tip.
  • Inspect tip cone for damage.
  • Clean the tip cone and shaft.
  • If problem persists, contact Gilson support.
Plunger movement is stiff or sticky
  • Brak smarowania
  • Contamination in the plunger mechanism
  • Consult the full manufacturer's manual for lubrication instructions.
  • Disassemble and clean the plunger assembly if comfortable, or send for professional servicing.

For issues not listed here or if problems persist, refer to the comprehensive Gilson Pipetman F100 user manual or contact Gilson technical support.

6. Specyfikacje

FunkcjaSzczegół
ModelPipetman F100
Zakres głośności100 µL (Fixed)
Typ pipetyAir-displacement, Fixed-Volume
ProducentGilson
Wymiary opakowania12.5 x 4.3 x 1.5 cala (31.75 x 10.92 x 3.81 cm)
Waga paczki8 uncji (0.5 funta)
ASINB008TPWFLA

7. Gwarancja i wsparcie

For detailed warranty information, service, or technical support regarding your Gilson Pipetman F100, please refer to the official Gilson website or contact Gilson customer service directly. Ensure you have your product model number (F100) and any purchase details available when seeking support.

Visit Gilson's Official Webstrona

© 2023 Gilson. All rights reserved. This manual is for informational purposes only.